Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add filters








Language
Year range
1.
Chinese Journal of Medical Instrumentation ; (6): 68-70, 2014.
Article in Chinese | WPRIM | ID: wpr-259931

ABSTRACT

This paper selects the bag-type infusion sets for single use as samples, which are produced by different manufacturers and based on the ethylene oxide sterilization. The ethylene oxide sterilization residues in different parts of samples are detected by colorimetric analysis. Combined the comparison of the ethylene oxide residues testing results in the different parts of the same sample with the actual situation in clinical use, more reasonable sampling positions are found to detect the ethylene oxide sterilization residues. The result of this experiment will play a guiding role in the detection of the actual samples.


Subject(s)
Disposable Equipment , Equipment Contamination , Ethylene Oxide
SELECTION OF CITATIONS
SEARCH DETAIL